在PEO法声速测量中确定反射相移时延修正值与判别正确重合的方法

DETERMINATION OF THE REFLECTION TIME DELAY AND A CRITERION FOR CORRECT OVERLAP IN ULTRASONIC WAVE VELOCITY MEASUREMENT WITH PEO METHOD

  • 摘要: 脉冲回波重合法(PEO法)目前已被公认为精密测定声速绝对值的最适用方案。为了使该法所具有的高精度测时能力起到实际效果,必须把其中所包含的反射时延、衍射时延等修正值相应地推求出来,加以修正。至今所广泛采用的反射时延计算方法,在实用上仍有所不便,在理论上也不尽准确,为此,本文进一步作了分析。
    本文得出了包括衍射影响在内的反射相移时延修正值的计算方法。并提出了一种双晶片法实验测定修正值的方案。本文并提出取零相移周期增量Δг,作为实验常数,其值在一般情况下约为1.7ns,这样对实用就更为简便。
    本文还提出了一种判别准确重合的方案。当准确重合时,测得声时一周期曲线的斜率为—h

     

    Abstract: The Pulse-Echo-Overlap method (PEO method) has now been considered to be the mest adequate method for accurate measurement of the absolute value of ultrasonic wave velocity. For reaching a realizable resoult of its high ablity of accurate time measurement, it is essential to find and implement its delay time corrections, e.g. reflection time delay, diffraction time delay, etc. Because of the unconvenience in practice of the widely used calculation method of reflection time delay, and its imperfection in theory, this paper makes a further investigation.
    In this paper, a calculating procedure of the reflection time delay correction is worked out, and is also suggested an experimental determination method for its value by means of double crystals. This paper further suggests a parameter of period increment of zero phase shift △τr, and considers it to be an experimental constant, whose value is about 1.7 ns in p- raetical applications.
    Furthermore, this paper suggests a criterion for correct overlap. If it is correctly over-lapped, the slope of the curve of ultrason isc wave trap time to be measured against RF period will equal to - h.

     

/